The Poet IQ2® Anesthetic Gas Monitor - medicalequipment

Post Top Ad

Responsive Ads Here

30 June 2011

The Poet IQ2® Anesthetic Gas Monitor


Powerful and compact—fits into small shelves on gas kits and is an excellent choice for updating existing anesthesia delivery systems with state-of-the-art agent detection and proprietary 5-agent gas analysis technology.

The Poet IQ2® Anesthetic Gas Monitor provides an affordable solution for 5-agent gas monitoring in any operating environment. The Poet IQ2 has an external video output port that is compatible with VGA flat screen displays, allowing patient data to be displayed where the surgical team needs to see the information. The anesthetic monitor comes with an internal high resolution color display that is designed into a compact, easy to use control panel that can fit into tight spaces. The internal display controls can be placed within easy reach while an external large screen display can be mounted where ever desired.

The Poet IQ2 automatically identifies and quantifies inspired and expired CO2, N2O, O2, and five anesthetic agents. The exceptional dynamic range of the advanced High IQ™ agent detector allows the monitor to accurately measure and display mixed gas measurements at low volumes.

The system's reliable performance, ease of use, flexible design, and affordable cost make it the ideal monitoring solution for anesthesia applications in hospitals and surgical centers.
State-of-the-art NDIR technology identifies and measures five anesthetic agent gases:

Halothane, Isoflurane, Enflurane, Sevoflurane, and Desflurane.
Automatic and manual primary agent detection.
Breath-by-breath CO2, O2, N2O monitoring.
Gas flow sampled at 100, 150, and 200 ml/min.
Mixed agent identification.
Measurement accuracy is not affected by alcohol or metabolic ketones.
Fast warm-up time ensures full accuracy within minutes.
No user calibration required; minimal user interaction.
Intuitive, easily configured user interface minimizes training and is ideal for multiple users.
Bright, real-time display of numerical values and waveforms provides instant notification of changing patient status.
Lightweight, portable design provides flexible workspace options.
Gas Monitoring
Method: Sidestream; Non-dispersive infra-red (NDIR)
Identified Gases: Halothane, Isoflurane, Enflurane,
Desflurane, Sevoflurane, CO2, N2O, Oxygen
Concentration Units: Vol%, Torr, kPa, mmHg
Flow Rates: 100 ml/min, 150 ml/min, 200 ml/min
Occlusion Clearing: Automatic
Auto Zeroing: Occurs 30 to 60 minutes
Duration: 3.0 to 7.0 seconds
Manual user calibration not required.
TemPerature stabilized optical assembly.

Agent Detection
Measurement Range:
Halothane: 0 to 10%
Isoflurane: 0 to 10%
Enflurane: 0 to 10%
Desflurane: 0 to 20%
Sevoflurane: 0 to 10%
CO2: 0 to 12.5%
N2O: 0 to 99%
Oxygen: 0 to 100%

Measurement Accuracy:
Agents: ± (0.1% abs. + 4% reading)
CO2: ± 0.2% abs. or 4% of reading
N2O: ± (1.5% abs. + 4% of reading)
Oxygen: ± 3 vol% (for 0-90%), ± 4 vol% (for 91-99%)

Time to Detect Agent: < 15 seconds @ 200ml/min
Agent Detection Resolution: 0.1 Volume Percent
Mixed Gas Threshold: 0.2 vol. % +10% of total concentrations

Rise Time:
Agents: 450 msec
CO2: 350 msec; N2O: 400 msec; Oxygen: 600 msec

Respiration Range
Range: 1 to 60bpm
Accuracy: ± 2bpm

System Features
Display
Main Display: Active color TFT
Size: 5.9" Display Area
Resolution: Internal Display; 320 x 240 pixels
External Video; 640 x 480 pixels
Number of Traces: 3 maximum

Controls
Keys: 7; membrane-activated
Rotary knob: Push and rotate; 24 steps/turn

System Outputs
Com 2; Serial/Analog, mini-DIN8
Com Ports: Com 1; RS 232-compatible, Serial DB-9
Video Port: Color VGA compatible, High Density D-Sub 15

Trends
Memory: 24 hours of tabular & graphic patient data

Power Requirements
Voltage: 100, 120, 220, 240 VAC; 50/60 Hz
Consumption: 40 W, typical

Mechanical
Weight: 13 lbs. (5.9 kg.)
Size: 6.5” (H) x 11.0” (W) x 12.0” (D);
16.5cm (H) x 27.94cm (W) x 30.48cm (D)

Environmental
Operating Temperature: 15º to 35º C; 59º to 95º F
Storage Temperature: -5º to 50º C; 23º to 122º F
Operating/Storage Humidity: 15% to 95%, noncondensing
Altitude: -1,000 to 10,000 ft. (-300m to 3000m)
Type of Protection: Class I Equipment
Degree of Protection: Type CF
Protection against ingress: Ordinary
Compatible with external VGA displays.
External display not included.
Specifications subject to change without notice.
GAS MONITORING ACCESSORIES

Sampling Devices Catalog Number Description
625N Sample Lines (Box of 25)
624 Disposable Nasal Cannula (Box of 10)
628 Divided Nasal Cannula (Box of 10)
616 Endotracheal Adaptor, Straight (Box of 10)
617 Endotracheal Adaptor, Elbow (Box of 10)

938F-NC WaterChek™ 2+ Water Trap (Box of 30)
644 Replacement O2 Cell
655 Scavenging Kit (Exhaust Line & Adaptor)
Calibration Accessories Catalog Number Description
621 Cal Gas Cylinder (5% CO2, Balance N2)
622 Cal Gas Cylinder (10% CO2, Balance N2)
716 Cal Gas Cylinder (95% O2, Balance CO2)
620 Cal Gas Cylinder (100% N2)
623 Hardware Regulator For Cylinder
613 Regulator Tubing
641 Cal Gas Aerosol (5% CO2, Bal N2)
642 Cal Gas Aerosol (10% CO2, Bal N2)
631 Cal Gas Aerosol (1% Isoflurane, 5% CO2, 60% N2O, Bal N2)
634 Cal Gas Aerosol (1% Enflurane, 5% CO2, 60% N2O, Bal N2)
635
Cal Gas Aerosol (1% Halothane, 5% CO2, 60% N2O, Bal N2)
1367 Cal Gas Aerosol (1% Sevoflurane, 5% CO2, 60% N2O, Bal N2)
1366 Cal Gas Aerosol (4% Desflurane, 5% CO2, 60% N2O, Bal N2)

Calibration Kits Catalog Number Description
612 CO2 only (Non-Anesthetic Agent)
5% CO2 cylinder, 10% CO2 cylinder, hardware regulator with
tubing, plastic storage case.
715 CO2 and O2 (Non-Anesthetic Agent)
5% CO2 cylinder, 10% CO2 cylinder, 95% O2 cylinder regulator
with tubing, plastic storage case.
1365 Anesthetic Agent Verification Kit (5 Agent)
One of each five aerosol agents, ISO, ENF, HAL, SEV, DES
as described above, includes plastic regulators with tubing.
909 Power Cord (Domestic US)
910UK Power Cord (United Kingdom)
910INT Power Cord (International)
910 Unterm Power Cord (Unterminated)
603 Japan Power Cord (Japan)

Post Bottom Ad