PORTABLE DOPPLER: ULTRASONIC FLOWMETER, FD613 - medicalequipment

Post Top Ad

Responsive Ads Here

29 July 2011

PORTABLE DOPPLER: ULTRASONIC FLOWMETER, FD613




The FD610 Series ultrasonic flow meter is made in USA and advertised at www.omega.com. It has features advanced Trans-Phase measuring technology, providing accurate and reliable flow velocity assessments in closed piping systems. The FD610 Series utilizes a non-invasive clamp-on transducer which is placed on the outside of a pipe. Within seconds, the large 18 mm (0.7") LCD provides stable flow rate readings in feet per second or meters per second. This product operates on metal or plastic pipes containing liquids with more than 100 ppm of 100 micron or larger suspended solids or entrained gases.



PRINCIPLE OF OPERATION
The FD610 Series flow meter utilizes an advanced Trans-Phase measuring technique, providing accurate, non-invasive fluid velocity assessment without opening the pipe. The FD610 Series utilizes two piezoelectric crystals contained within one transducer to transmit ultrasonic energy into the fluid stream and receive back energy reflected off discontinuities (suspended particles or entrained gases) within the moving liquid. Transformations that result from the energy reflections are processed and converted to a measurement of fluid velocity by the sophisticated software algorithm. The processor also controls all operations of the instrument from its ultra-efficient battery management circuitry to a proprietary FIR (Finite Impulse Response) filtration program.


SPECIFICATIONS
Supply Voltage:
Battery powered, non-rechargeable alkaline, four AA cells, providing over 30 hours of continuous operation

Flow Range:
0.30 to 30.00 FPS (0.10 to 9.00 MPS)

Display:
Single line – four-digit LCD readout of velocity (18 mm (0.7") digits), signal strength, measuring units, and low battery indication

Temperature:
-20 to 60°C (-28 to 140°F)

Enclosure Rating/Dimensions:
NEMA-12X ABS plastic; 200 H x 100 W x 38 mm H (7.76 x 4 x 1.5")

Weight: 0.7 kg (1.5 lb)

Accuracy: ±2% full scale

Transducer
Transducer Material:
Plated body; ULTEM 1000 sensor material

Cable: 2 m (6.6 ft) cable and connector

Temperature: -40 to 82°C (-40 to 180°F)

Humidity: 0 to 95% non-condensing
Mounting Method:
Clamp-on style with DOW 111 coupling compound